    Technical Specifications NIBP Measure Method: Automatic oscillometric Patient: Adult and Pediatric Measurement unite: mmHg/kPa selectable Measurement mode: Manual, Auto, STAT Measurement Type: SYS, DIA, MAP Measurement interval in AUTO mode:1,2,3,5,10,15,20,30,45,60Min Measuring period in STAT mode: 5 Min Measure range: Adult:30~2550 mmHg; child: 30~160 mmHg Accuracy: 1 mmHg Alarm limits range: SYS : Upper limits70~250 mmHg, Lower limits 60~180 mmHg DIA : Upper limits 40~200 mmHg, Lower limits35~150 mmHg MAP : Upper limits 60~230 mmHg, Lower limits 50~160 mmHg SpO2 Measure Range: 35~100% SpO2 Accuracy: 2%(70~99%), 3%(40~70%) Pulse Rate Range: 0~254 bpm Pulse Rate Accuracy: 1 bpm Alarm limits range: SpO2 upper limits : 95~100% lower limits: 85~96% Pulse Rate:keep up with the heart rate Temperature Channel: 1 Input: Body surface thermal-sensitive resistor temperature sensor Measurement Unit: °C/°F selectable Measure range: 0~50℃ Alarm limits range: Upper limits 36~45℃ Lower limits 34~38℃ Measure accuracy: 0.1℃ System specifications Alarm ON/OFF: controlled by general switch and parameter switch (two levels) Display mode: LED and 2.8’ color LCD screen Trend data: 3888 groups of SpO2, TEMP,SYS, DIA trend data, The record time interval can be set to 5 Sec. to 120 Sec. , so that the record time rang can be up to 128 hr. AC power adaptor Input: AC 100~240V, 50/60Hz, 0.7A Output: DC 5V,2.5A Internal battery Type: 2800mAh/3.7V Li-ion rechargeable battery Operating Time: 6 hours of stand-by time with a full battery. Recharge time: 8 hours Dimension:280mm×190mm×100mm Net Weight : 1000g

   
 